Download Catalog
Micro-Spectrophotometer BZPS-502 offers fast, precise analysis of nucleic acids, proteins, and cellular samples using minimal volume. It employs four-path optical detection for superior stability, linearity, and reproducibility across a broad range. This device's fluorescence function enables ultra-sensitive quantification of DNA using compatible fluorescent reagents. Its built-in printer allows easy report generation directly from the device for streamlined workflows. Our Micro-Spectrophotometer BZPS-502 with a xenon flash lamp and low light stimulation, ensures quick detection and extended light source lifespan.
| Wavelength Range | 190 nm to 850 nm |
|---|---|
| Optical Path | 0.01 mm, 0.02 mm, 0.05 mm (High concentration measurement), 0.2 mm, 1.0 mm (General concentration measurement) |
| Number of Detection Channels | 1 |
| Detector | HAMAMATSU UV enhanced CMOS linear array sensing |
| Nucleic Acid Detection Range | 0 to 27,500 ng/uL dsDNA |
| Protein Detection Range | 0.06 to 820 mg/mL BSA |
| OD 600 Absorbance Range | 0 to 6.000 Abs |
| OD 600 Absorbance Stability | (0,3) ≤ 0.5%, (3,4) ≤ 2% |
| OD 600 Absorbance Accuracy | (0,2) ≤ 0.005 A, (2,3) ≤ 1%, (3,4) ≤ 2% |
| OD 600 Absorbance Repeatability | (0,3) ≤ 0.5%, (3,4) ≤ 2% |
| Sample Volume Requirement | 0.5 to 2 ul |
| Light Absorption Range | 0.02 to 300A (Equivalent to 10 mm) |
| Absorbance Accuracy | 0.002 Abs (0.2mm optical path) |
| Spectral Resolution | < 1nm |
| Photometric Accuracy | ± 0.005 A or ± 1% |
| Wavelength Accuracy | ± 1 nm within |
| Wavelength Reproducibility | < ± 0.2 nm |
| Sample Wavelength Scanning | 190 to 850 nm |
| Base Measurement | 0.5 to 2ul |
| Detection Time | < 6 sec. |
| Sample Base Material | Quartz Fiber and Stainless Steel |
| Illuminant | Xenon flash lamp |
| Data Output Mode | USB |
| Printer | Yes |
| Standby Power Consumption | 5 W |
| Power Adapter | 24 V, DC |
| Power Dissipation | 25 W |
| Weight | 5 Kg |
| Dimension | 200 × 260 × 165 mm |
Four-Path Optical Detection
Fluorescence Quantification
OD600 Microbial Detection
Built-In Printing
Xenon Flash Lamp Source
The Micro-Spectrophotometer BZPS-502 is widely used for accurate quantification of DNA, RNA, and proteins in molecular biology and genomics. It enables convenient microbial concentration analysis in microbiology and biotechnology applications. Its fluorescence capability supports ultra-sensitive assays in clinical diagnostics and pharmaceutical research.